2051 CD/CG
- Series 2051 Pressure measurement - SPECIFICATIONS 2051 CD/CG
- Accuracy: ±0,075% of span (optional 0,065%)
- Stability: 2051C range 2-5 ±0,1% of URL for 2 years | (optional ±0,125% of URL for 5 years kód-P8) | 2051CD range 1 ±0,2% of URL for 1 roku
- Rangetown: 100:1
- Pressure range: CD: -13,79 MPa - 13,79 MPa | CG: -98 kPa - 13,79 MPa
- Number of ranges: 5
- Ambient temperature limit: -40°C to +85°C
- Temperature process fluids: -40°C to +121°C (coplanar collar, cartridge D.C. Silicone)
- Output: 4-20 mA + HART or Fieldbus
- Overpressure limit: range 1 13,79 MPa | range 2-5 25 MPa ( optional 31 MPa kód-P9)
- Certifications: Atex
- Data sheet - 2051